Abstract
The toxicity of sediments is often closely bound to the labile toxic elements contents. In porewaters,
metal concentrations are generally measured after separation from the solid material by means of pressure or
centrifugation. The recent developments of in situ metal measurements by microsensors do not require any
pretreatment of the sediment and hence avoid some possible artefacts like oxygen penetration in anoxic sediment
and/or temperature modification. Our work describes firstly the preparation of microelectrodes based on a thin solid
Silver (diameter: 30
m) or Iridium (diameter: 75
m) wire, covered at the tip with a fine layer of mercury.
Secondly, an analytical procedure (based on cathodic voltammetry) is presented to measure Mn(II), Fe(II) and S(-II)
concentrations, which play a relevant role in the metal pollutant cycle in sediments. By anodic stripping
voltammetry, such microsensors are able to detect trace heavy metals such as Cd(II) or Pb(II) at relatively low
concentration levels (~1
g.dm
-3) with 5 minutes accumulation time. Finally, the most important challenge is to
install the microelectrodes in natural media and to follow up metal pollution over a long time period.
© EDP Sciences 2003