Issue |
J. Phys. IV France
Volume 08, Number PR3, June 1998
Proceedings of the 3rd European Workshop on Low Temperature ElectronicsWOLTE 3 |
|
---|---|---|
Page(s) | Pr3-233 - Pr3-236 | |
DOI | https://doi.org/10.1051/jp4:1998351 |
Proceedings of the 3rd European Workshop on Low Temperature Electronics
WOLTE 3
J. Phys. IV France 08 (1998) Pr3-233-Pr3-236
DOI: 10.1051/jp4:1998351
Space Research Organization Netherlands, Sorbonnelaan 2, 3584 CA Utrecht, The Netherlands
© EDP Sciences 1998
WOLTE 3
J. Phys. IV France 08 (1998) Pr3-233-Pr3-236
DOI: 10.1051/jp4:1998351
Trade-off study of the SQUID read-out for a hot-electron micro-calorimeter
M. Frericks, H.F.C. Hoevers, P. de Groene, W.A. Mels and P.A.J. de KorteSpace Research Organization Netherlands, Sorbonnelaan 2, 3584 CA Utrecht, The Netherlands
Abstract
SQUIDs and their read-out electronics for application with a hot-electron bolometer are considered in detail. This results in a theoretical trade-off study between a Hypres two-stage DC-SQUID, APF, DROS and an Oxford Instruments' Internal Relaxation Oscillation Mode SQUID. The study shows that only a Hypres DC-SQUID can meet the requirements on noise, dynamic range and speed as derived from the application for an X-ray micro-calorimeter. The scope for redesign of the other types of SQUID is addressed.
© EDP Sciences 1998